TDS Calculator (Total Dissolved Solids)
Convert conductivity to TDS, TDS to conductivity, or work out TDS from a weighed residue, and see exactly how the answer is built. The big question with any TDS conversion is which factor (k) to use, so this calculator gives you a water-type picker with the reasoning behind each choice, shows how much the choice changes your result, and can derive your own factor from a lab measurement. It also decodes handheld meters that show “ppm” on a 500 or 700 scale, and corrects readings for temperature.
Background
Total dissolved solids is the mass of everything dissolved in a litre of water (mostly mineral salts), reported in mg/L, which is practically the same as ppm for dilute water. The reference measurement is gravimetric: filter, evaporate and weigh the residue. Because that takes hours, TDS is usually estimated from electrical conductivity: TDS = k × EC, with EC in µS/cm at 25 °C. The factor k is not a constant of nature. It depends on which ions are dissolved: about 0.5 for sodium chloride, 0.55 to 0.70 for most natural waters, and up to about 0.9 for sulfate-rich or very salty water.
How to use this calculator
- Choose EC → TDS when you have a conductivity reading. Pick the kind of water, and you get the TDS, a plausible range, a table showing what every water type would give for the same EC, and where the result falls on the taste scale.
- Choose TDS → EC to find the conductivity you should see for a given TDS, for example when mixing a test solution or checking whether a lab result is plausible.
- Choose Gravimetric to turn dish and residue weights into TDS. It checks that the residue is inside the 2.5–200 mg window and tells you what volume to take next time.
- Choose Find My k when you have a lab TDS and an EC for the same water. It works out k for each pair and the mean, and flags pairs outside the usual 0.55–0.70 band.
- Choose Meter Scale when a handheld meter shows “ppm” and you are not sure what that means. It recovers the real conductivity, shows the reading on other scales, and estimates the actual dissolved mass.
- Turn on temperature correction if your meter does not compensate automatically and the sample is not at 25 °C. Then click Calculate to see the result, a callout explaining what it means, a diagram for that task and every step with your numbers substituted.
How to choose the conversion factor (k)
The factor links conductivity to mass. It differs because ions differ in how much current they carry per milligram: small, highly mobile ions such as sodium and chloride carry a lot (so the same EC means less mass, a low k), while calcium, magnesium and sulfate carry less (so the same EC means more mass, a high k). The table gives starting values; the numbers come from Standard Methods 2510 and 1030, lab data on NaCl solutions, and an extension-service table for natural waters.
| Water type | Typical k | Range | Choose it when |
|---|---|---|---|
| Natural fresh water | 0.65 | 0.55 to 0.70 | Tap water, rivers, lakes and most wells with EC from about 100 to 1,500 µS/cm. The safe default. |
| Natural, scaled to EC | 0.50 to 0.82 | ± 0.05 allowance | EC well outside 100 to 1,500 µS/cm. Natural water gets a larger k as it gets saltier. |
| NaCl-dominated | 0.50 | 0.48 to 0.60 | RO permeate, softened water, road-salt runoff, anything where sodium chloride is the main salt. |
| Hard / sulfate-rich | 0.80 | 0.70 to 0.90 | Gypsum-rich groundwater, some mine and industrial waters. |
| Brackish / saline | 0.75 | 0.70 to 0.85 | EC above about 5,000 µS/cm, estuaries, seawater influence. |
| Meter 500 scale | 0.50 | fixed | You want to reproduce what a meter on its 0.5 setting shows. |
| Meter 700 scale | 0.70 | fixed | You want to reproduce what a meter on its 0.7 setting shows (common in hydroponics). |
Best: measure it. Evaporate one sample (Gravimetric mode), take the EC of the same water, and let Find My k divide them. Two or three pairs across your normal range are better than one.
If you know the chemistry: a lab ion analysis, or a known source (RO system, gypsum aquifer, coastal well), points straight to a row in the table.
If you know nothing else: use natural fresh water (0.65) when EC is moderate, and report the range. If EC is very low or very high, use the EC-scaled option instead.
If you are matching a meter: use its own scale (0.5 or 0.7) so your numbers agree with the display, but remember that is a convention, not the true dissolved mass.
Sanity check: a measured TDS ÷ EC outside 0.55 to 0.70 is a prompt, under Standard Methods, to recheck the analysis. Waters that are very acidic or contain strong hydroxide can have lower factors, and very saline waters higher ones.
How the conversion works
Conductivity measures how well water carries electric current, in µS/cm. Dissolved ions carry the current, so more ions mean higher EC. Pure water has an EC of only about 0.055 µS/cm.
EC changes with temperature, about 1.9% per °C for typical water, so values are referenced to 25 °C. A reading taken at 18 °C is lower than the 25 °C value and must be scaled up before using k.
TDS is then estimated as k × EC25. The result is an estimate whose uncertainty is mostly the uncertainty in k, which is why this calculator shows a range instead of a single value when the factor is uncertain.
Gravimetric TDS is measured directly: filter through a glass-fibre filter, evaporate the filtrate in a weighed dish, dry to constant weight at 180 °C, and weigh. Dividing the residue mass by the volume gives mg/L.
Handheld “TDS meters” are conductivity meters with a built-in multiplier, usually 0.5 or 0.7. Two meters on different scales read the same water differently, so always note the scale with the number.
TDS does not identify what is dissolved. Harmless minerals and harmful contaminants both count, and substances that do not ionise (sugar, many organic compounds) barely change EC. WHO sets no health-based guideline value for TDS, only taste ratings.
Formulas & Equations Used
Unit conversion: 1 mS/cm = 1 dS/m = 1,000 µS/cm; 1 S/m = 10,000 µS/cm
Temperature to 25 °C: EC₂₅ = EC(T) ÷ (1 + α × (T − 25)), α ≈ 0.019 per °C
EC to TDS: TDS (mg/L) = k × EC₂₅ (µS/cm)
TDS to EC: EC₂₅ = TDS ÷ k, EC(T) = EC₂₅ × (1 + α × (T − 25))
Range from an uncertain k: TDS between k_low × EC₂₅ and k_high × EC₂₅
Your own factor: k = lab TDS (mg/L) ÷ EC₂₅ (µS/cm), mean k = Σk ÷ n
Gravimetric TDS: TDS (mg/L) = (A − B) × 1000 ÷ V(mL) = residue (mg) ÷ V (L)
Sample volume to target a residue: V (L) = target residue (mg) ÷ expected TDS (mg/L)
Meter scale to EC: EC₂₅ = meter reading ÷ k_meter, other-scale reading = EC₂₅ × k_other
Between two scales: reading₇₀₀ = reading₅₀₀ × 0.7 ÷ 0.5 = 1.4 × reading₅₀₀
Example Problems & Step-by-Step Solutions
Example 1 — Tap water, EC to TDS
Tap water reads 450 µS/cm at 25 °C. Estimate the TDS.
Step: Natural fresh water, k = 0.65 (range 0.55 to 0.70). TDS = 0.65 × 450 = 292.5. Low end 0.55 × 450 = 247.5; high end 0.70 × 450 = 315.
Result: about 292.5 mg/L, plausibly 247.5 to 315 mg/L, in the WHO “excellent” taste band (under 300).
Example 2 — A cold sample
A meter without compensation reads 820 µS/cm in 18 °C river water. What is the TDS?
Step: EC₂₅ = 820 ÷ (1 + 0.019 × (18 − 25)) = 820 ÷ 0.867 = 945.79 µS/cm. TDS = 0.65 × 945.79 = 614.76. Without the correction: 0.65 × 820 = 533.
Result: 614.76 mg/L. Skipping the temperature correction would have under-reported by about 13%.
Example 3 — Brackish water
A coastal well measures 8.5 mS/cm. Estimate its TDS.
Step: 8.5 mS/cm = 8,500 µS/cm. Saline factor k = 0.75 (range 0.70 to 0.85). TDS = 0.75 × 8,500 = 6,375. Range: 0.70 × 8,500 = 5,950 up to 0.85 × 8,500 = 7,225.
Result: about 6,375 mg/L, between 5,950 and 7,225 mg/L. This is far above the 1,200 mg/L taste threshold.
Example 4 — TDS to EC
You want test water at 500 mg/L TDS. What EC should a meter show?
Step: EC₂₅ = TDS ÷ k = 500 ÷ 0.65 = 769.23 µS/cm. A larger k gives a lower EC: 500 ÷ 0.70 = 714.29. A smaller k gives a higher EC: 500 ÷ 0.55 = 909.09.
Result: aim for about 769 µS/cm, expecting anywhere from 714 to 909 µS/cm depending on the salts.
Example 5 — Gravimetric TDS
A dish weighs 52,341.2 mg empty and 52,372.4 mg after 100 mL of filtered water is evaporated and dried at 180 °C. Find the TDS.
Step: Residue = 52,372.4 − 52,341.2 = 31.2 mg. Volume = 0.100 L. TDS = 31.2 ÷ 0.100 = 312 mg/L. To leave about 100 mg next time: 100 ÷ 312 = 0.32 L, about 320 mL.
Result: 312 mg/L. The 31.2 mg residue is inside the 2.5 to 200 mg window, so the weighing is reliable.
Example 6 — Find your own k
Four samples give lab TDS and EC of 310 and 480, 420 and 640, 560 and 810, 905 and 1,300. What factor should you use?
Step: k = 310÷480 = 0.646, 420÷640 = 0.656, 560÷810 = 0.691, 905÷1,300 = 0.696. Mean = (0.646 + 0.656 + 0.691 + 0.696) ÷ 4 = 0.672.
Result: use k = 0.672. All four values are inside 0.55 to 0.70, so the data look consistent. Note that k rises slightly with EC.
Example 7 — What does my meter's ppm mean?
A pen meter on its 500 scale shows 350 ppm. What would a 700-scale meter show, and what is the real TDS of natural water?
Step: EC = 350 ÷ 0.5 = 700 µS/cm. On the 700 scale: 0.7 × 700 = 490 ppm. Real TDS for natural water: 0.65 × 700 = 455 mg/L (range 385 to 490).
Result: the same water reads 350 or 490 ppm depending on the scale, and holds roughly 455 mg/L of dissolved solids.
Example 8 — The cost of a wrong factor
A salty groundwater sample reads 3,000 µS/cm. How wrong is the flat 0.65 factor?
Step: Flat factor: 0.65 × 3,000 = 1,950 mg/L. Natural-water table value at 3,000 µS/cm is k = 0.75: 0.75 × 3,000 = 2,250 mg/L. Difference = 300 mg/L, which is 300 ÷ 2,250 = 13.3% of the better estimate.
Result: the flat 0.65 would under-estimate by about 300 mg/L. For moderate EC the error is small; for salty water it is not.
Frequently Asked Questions
What is TDS?
Total dissolved solids is the combined mass of all inorganic salts and small amounts of organic matter dissolved in a litre of water, usually reported in mg/L. It mostly consists of calcium, magnesium, sodium, potassium, bicarbonate, chloride and sulfate.
How do I convert EC to TDS?
Multiply conductivity in µS/cm (at 25 °C) by a factor k: TDS (mg/L) = k × EC. For most natural waters k is between 0.55 and 0.70, with 0.65 a common middle value. If your EC is in mS/cm, multiply by 1,000 first.
Which conversion factor should I use?
If you can, measure it: take a lab TDS and an EC for the same water and divide them (Find My k mode). Otherwise pick by water type: 0.65 for typical fresh water, about 0.5 where sodium chloride dominates, 0.7 to 0.9 for sulfate-rich or saline water. See the decision guide above.
Why does k change from water to water?
Each ion carries a different amount of current per milligram. Sodium and chloride are light and mobile, so they give a lot of EC per mg (low k). Calcium, bicarbonate and sulfate give less, so the same EC represents more mass (higher k). The mix of ions therefore sets k.
What are the 500, 442 and 700 scales?
They are meter conventions. The 500 scale uses k = 0.5 and is based on sodium chloride. The 442 scale is a mixed-salt “natural water” standard from Myron L, and the 700 scale (k = 0.7) is used by many hydroponic meters; the 442 and 700 conventions behave similarly. Meters often let you choose between 0.5 and 0.7.
Is mg/L the same as ppm?
For dilute water (density near 1 g/mL), yes: 1 mg/L ≈ 1 ppm by mass. For very salty water the density is noticeably above 1 and the two drift apart slightly.
Why does temperature matter?
Warmer water carries current better: EC rises roughly 1.9% per °C. Standard values are therefore referenced to 25 °C. Most modern meters compensate automatically; if yours does not, correct with EC₂₅ = EC ÷ (1 + 0.019 × (T − 25)).
What is the most accurate way to measure TDS?
Gravimetric measurement: filter, evaporate and weigh the residue (Standard Methods 2540 C). It measures the mass directly, with no assumption about which ions are present. It is slow, needs a balance and an oven, and works best when the residue is 2.5 to 200 mg.
Why dry at 180 °C instead of 103–105 °C?
At 103–105 °C the residue can keep water of crystallisation and trapped water, which makes the result read higher. At 180 °C almost all of it is driven off, and the result is closer to the sum of the individual dissolved minerals. Compare results only if they were dried at the same temperature.
How much water should I evaporate?
Choose a volume that leaves 2.5 to 200 mg of dried residue. Less than 2.5 mg and balance error dominates; more than 200 mg can form a crust that traps water. The Gravimetric mode tells you the volume range for your water.
What if my measured k is outside 0.55–0.70?
Standard Methods treats a measured TDS ÷ EC outside 0.55 to 0.70 as a signal to recheck the analysis. Real exceptions exist: very acidic or strongly alkaline water can fall below, and very saline or sulfate-rich water above. Check the weighing, the filter and the EC calibration first.
Does a low TDS mean the water is safe?
No. TDS measures total dissolved material, not what it is. WHO sets no health-based guideline for TDS; its ratings are about taste. Harmful dissolved substances such as lead, arsenic or nitrate can be present in low-TDS water, and RO water has very low TDS by design.
What TDS is good for drinking water?
WHO taste panels rated water under 300 mg/L excellent, 300 to 600 good, 600 to 900 fair, 900 to 1,200 poor and above 1,200 unacceptable. The US EPA has a non-enforceable secondary standard of 500 mg/L. Extremely low TDS water can taste flat.
Why does my meter disagree with the lab?
The usual causes are a different factor (meter on 0.5, lab by evaporation), an uncompensated temperature, a miscalibrated cell, or water whose ions do not suit the meter's built-in scale. Convert the meter reading back to EC in Meter Scale mode, then compare like with like.
Can I use this for hydroponics or aquariums?
Yes for the conversions. Hydroponic targets are usually quoted as EC in mS/cm, or as “ppm” with the scale named (500 or 700). Use Meter Scale mode to translate between them. For aquariums, track EC or TDS as a trend; the exact mg/L depends on the salts you use.
Can I estimate TDS for seawater?
Roughly. Seawater at about 53,000 µS/cm holds around 35,000 mg/L, a ratio of roughly 0.7 (published values run from about 0.65 to 0.75). For salinity work, practical salinity scales are better than a TDS factor, and this calculator is not intended for ocean-science accuracy.
Where the numbers come from
- Standard Methods for the Examination of Water and Wastewater, 2510 (conductivity) and 1030 (checking analyses): the 0.55 to 0.70 range for natural waters, factors up to about 0.9 for saline or sulfate-rich water, and the TDS ÷ EC consistency check.
- Standard Methods 2540 B and C (solids): the 2.5 to 200 mg residue window, the 180 °C drying temperature for TDS, and the formula (A − B) × 1000 ÷ sample volume.
- WHO, Total dissolved solids in drinking-water (background document): the taste ratings and the statement that no health-based value is set.
- North Dakota State University Extension, Using electrical conductivity and TDS meters to field test water quality: the table of factors for natural water that rise from 0.50 at 100 µS/cm to 0.82 at 10,000 µS/cm.
- Manufacturer documentation (for example Hanna Instruments) for the 0.5 and 0.7 meter scales and the 1.9% per °C compensation coefficient. Typical values are rules of thumb, and the right factor for your water is the one you measure.