Skip to main content
뒤로

Microscopy and Cytology - General Biology lab 5

컨트롤 버튼이 '내비게이션' 모드로 변경되었습니다.
1/20
  • Function of the arm on a compound microscope

    Used to carry the microscope; located between the body tube and the base.
  • Purpose of the condenser in a compound microscope

    Substage lens that concentrates light on the specimen; its position can be adjusted to vary light intensity.
  • What is total magnification in microscopy?

    The product of the ocular lens magnification and the objective lens magnification.
  • Magnification of the ocular lens in the lab microscopes

    10X; the eyepiece magnifies the image ten times.
  • Magnifications of the four objective lenses used

    Scanning: 4X, Low power: 10X, High power: 40X, Oil immersion: 100X.
  • How to calculate total magnification with a 40X objective lens

    Total magnification = 10X (ocular) × 40X (objective) = \(400X\).
  • Definition of resolution in microscopy

    Resolving power is the ability to distinguish two close objects as separate; compound microscopes resolve objects about 0.2 µm apart.
  • What is parfocal in compound microscopes?

    Once focused on low power, the image remains nearly in focus when switching to higher magnifications.
  • Proper use of coarse and fine focus knobs

    Use coarse focus only on low power; use fine focus for final adjustments and on high power.
  • Function of the iris diaphragm

    Regulates the amount of light passing through the stage to improve image contrast.
  • What is the field of view in microscopy?

    The circular area visible through the microscope lens; its diameter decreases with increasing magnification.
  • What is depth of field in microscopy?

    The vertical distance in focus; decreases as magnification increases.
  • How to prepare a wet mount slide

    Place specimen on slide, add liquid (water or stain), cover with coverslip avoiding air bubbles, then observe under microscope.
  • Purpose of staining specimens

    Stains absorb light and improve contrast to reveal cellular structures.
  • Ocular micrometer calibration at low power (10X objective)

    One ocular unit equals \(10\ \mu m\).
  • Ocular micrometer calibration at high power (40X objective)

    One ocular unit equals \(2.5\ \mu m\).
  • How to measure specimen size using an ocular micrometer

    Count ocular units spanning the specimen and multiply by the ocular unit length for the objective used.
  • Difference between red and white blood cells under microscope

    Red blood cells are small, round, pinkish, and lack a nucleus; white blood cells are larger with dark-stained nuclei.
  • Why never use coarse focus on high power

    Coarse focus can damage the slide or lens because the objective is very close to the specimen.
  • How to carry a microscope safely

    Use both hands: one on the arm and one under the base; keep the cord secure to avoid tripping.